On this page
Signal amplitude at the selected terminals
A simulator can show a selected signal amplitude while a reference instrument reads a particular pair of outputs. Name that pair on the request and state the units and rate settings used by your engineering team. If one service lead gives an intermittent trace, identify it by tag or connector and keep its observation separate from the simulator’s output measurements.
ECG simulator as a reference instrument
List the generated functions, waveform settings and output connections that the department actually uses. Its calibration request differs from testing a monitor with that simulator. RTS calibrates the specified simulator outputs, recording their waveform settings and reference conditions.
Specifications and service checks
A lead that works on one output connector may give an intermittent trace through an adapter. List the simulator terminals and adapter used with the engineering test lead so the job can distinguish the output from the connection.
- Range and configuration
- The published heart-rate span includes 20–300 beats per minute. Rate, waveform shape and signal amplitude have separate tolerances.
- Manufacturers and models
- Rigel’s UNI-SIM includes an ECG and rhythm simulation function.
- What we check
- Measure output timing and amplitude with independent references at the selected outputs and settings. Name the waveform and connection arrangement on the certificate.
- Reference equipment and accuracy
- Use a traceable timebase and a calibrated voltage measurement system with enough low-level resolution and input impedance for the signal.
- When to repeat the check
- Review drift in the simulator’s amplitude and timebase when setting recall. Recheck after output circuit or connector repairs.
Technical references Rigel Medical product reference · Rigel UNI-SIM technical specifications