Electrical safety analyzer testing

An electrical safety analyser may measure protective-earth resistance and leakage current on the same bench. These functions use different connections and units. An order that simply asks for an electrical check may leave out the measurement network the hospital uses most.

Send an enquiry
Electrical safety analyzer — equipment illustration
Equipment illustration
On this page

Leads, test current and analyser mode

List the analyser modes, leads and input configuration shown on your service worksheets. For resistance, identify the test current setting where the instrument offers a choice. For current, include the measuring network or mode so comparisons use the same part of the analyser.

A service programme may call an observation ‘leakage’ while a saved result file lists several leakage modes. Include a blank copy of the worksheet or a report with personal details removed and mark the field your engineers use. It ties the analyser function and accessory leads to the term in your records.

Electrical safety analyzer calibration

Give the resistance, leakage-current, voltage and any simulated functions used, with connection modes and accessories. List each analyzer function and connection mode required by the workshop procedure. RTS calibrates those functions separately from testing the connected medical equipment.

Specifications and service checks

On an equipment register the analyser and its test leads may have separate asset numbers. If the leads are normally zeroed as a set, send their identities with the enquiry. A change in lead resistance can matter at the low end of a bond measurement.

Range and configuration
Protective conductor resistance and leakage current are different ranges. Record the chosen measuring network and test configuration as well as the displayed unit.
Manufacturers and models
Rigel’s 288+ is a portable electrical safety analyser for medical equipment. Fluke Biomedical’s ESA615 is another analyser for medical electrical safety measurements.
What we check
Compare resistance and leakage-current indications with suitable reference values. Verify leads and measuring network as part of the configuration when the procedure requires them.
Reference equipment and accuracy
Use calibrated low-resistance standards and a traceable leakage-current verification system rated for the analyser. A general multimeter alone cannot verify the network’s frequency response.
When to repeat the check
Check leads and adapters between calibrations. Reverify after an input fault or measuring-network repair.

Technical references Rigel Medical product reference · Fluke Biomedical product reference

Enquire about Electrical safety analyzer