On this page
Sensitivity and the paired level heads
Record the head numbers, cable arrangement, selected sensitivity and base length. State whether you need an absolute inclination reading from each head or the difference shown by the paired system. Include a normal working tilt on both sides of level.
For aligning a long bed, one head may remain on a reference surface while the other is moved from station to station. The record must identify the stationary head, since exchanging the two changes the sign and meaning of a differential reading.
Specifications and service checks
An alignment survey can join successive slope readings into a height profile. Keep the distance between stations with every reading. Identical slopes accumulated over 200 mm and 500 mm correspond to different changes in elevation.
- Range and configuration
- WYLER Clinotronic PLUS measures inclination up to ±60°. A high-sensitivity machine level may cover a much narrower span quoted in mm/m. Identify the display unit before choosing points.
- Manufacturers and models
- WYLER Clinotronic PLUS is a portable electronic inclinometer. A Leica Sprinter is instead a survey level used with a staff.
- What we check
- Calibrate the sensing head with its cable and readout at zero and at positive and negative slope. If two heads are used differentially, record their individual orientation and how the readout combines them.
- Reference equipment and accuracy
- Use a calibrated angle table, level comparator or sine setup capable of reproducing the selected small slopes. Include the uncertainty of the mounting surface.
- When to repeat the check
- Review zero in both orientations after transport. Recheck after changing the electronic head, base or compensation settings.
Technical references WYLER Release 2.0 · WYLER Digital level reference