+91 70565 55582[email protected]Get a quoteRequest a call
RTS Calibration
Menu

Electronic universal testing machine calibration

An electronic universal testing machine may switch between several load cells and extensometers while retaining the same frame and software. The serial number on the frame alone cannot identify the measuring channel that produced a tensile curve.

Send an enquiry
Electronic universal testing machine — equipment illustration
Equipment illustration
On this page

Force range for each fitted load cell

For each load cell, give its ID, working range, loading direction and the units selected in the software. Then identify the source of the displacement reading. Crosshead travel and extension at a specimen extensometer need separate checks if both enter the report.

When a low-capacity cell is fitted for a thin strip, the high-capacity cell used for steel bars may still be on the software list. The report should identify which cell was physically connected for each set of force readings.

Specifications and service checks

A software recipe can stop at a target crosshead movement while the extensometer measures strain directly on the specimen. If two displacement values disagree, name the channel used to control the run and the channel used to report extension.

Range and configuration
Instron’s 6800 range includes a 100 kN system. The fitted load cell can have a lower capacity than the frame, and a lower-capacity cell may be used for small specimens.
Manufacturers and models
Instron 6800 is a universal testing system family. Identify the particular load cell and whether it is used for tension, compression or both.
What we check
Verify the force indication through the selected electronic range and record the load cell and software channel. If the result depends on extension, verify the extensometer channel in its working configuration as well.
Reference equipment and accuracy
Use a calibrated force-proving device rated and classified for the working forces and direction. Use a displacement standard and time reference where travel or speed is included.
When to repeat the check
Review force and extension verification when a load cell, extensometer, grip train or drive component is changed. A check specimen helps reveal changes between calibrations.

Technical references Instron 6800 Series Universal Testing Systems · ISO Force-proving instrument calibration

Enquire about Electronic universal testing machine

Service enquiry