Preload and the transition through zero
Give the rated force, normal preload and whether the process crosses through zero from tension into compression. State whether you want the transducer output or the complete engineering-unit indication. Keep the cable and any inline amplifier identified if they are part of that measurement.
A controller receiving a voltage from an amplifier and a meter reading the raw bridge output do not share a numerical scale. If both are used, identify their units and ask for each reading at the same applied force for an interpretable comparison.
Specifications and service checks
A bridge rated at 2 mV/V gives a nominal 10 mV at full load with 5 V excitation. If the controller records volts after an amplifier, identify its gain as well as the transducer rating before comparing the two outputs.
- Range and configuration
- HBK’s U10M family includes nominal capacities from 1.25 kN to 1.25 MN. A load cell is chosen for one rated capacity, direction and mounting arrangement.
- Manufacturers and models
- HBK U10M is a tension/compression family. Compression-only designs such as the HBK C10 have different fittings.
- What we check
- Apply known forces in a sequence that shows repeatability, zero return and any difference on unloading. State whether the readout is part of the calibrated chain and report the directions tested.
- Reference equipment and accuracy
- For a transducer supplied without its readout, establish the bridge response at known forces and record excitation and output units. For an assembled chain, the master is a force standard and the result can be expressed directly in N or kN.
- When to repeat the check
- Overload, replacement of a readout or changes to load introduction justify another check. Keep the pre-load zero record for cells fitted permanently in a press.
Technical references HBK Compression load cell reference · ISO Force-proving instrument calibration