On this page
Where the gauge contacts the gap
Show the measuring feature and the surfaces on the part that it contacts. If the gauge carries clearly marked acceptance limits, include both values. For a special fixture with a handle, identify the asset marking on the working piece as well as the marking on the handle.
A gauge marked with a finished component number may have several gaps for different features. Number the gaps on a photograph and put the applicable drawing limit beside each one to keep results unambiguous.
Specifications and service checks
A tapered blade used to read a slot can enter farther if it is tilted. Identify the edge where the reading is taken and whether the inspection calls for a marked reading or simply acceptance between two fixed limits.
- Range and configuration
- Specify the intended gap and its unit. For a leaf set, the published RTS entry for feeler thickness covers 0.03 to 2 mm. A fixed gap fixture has its own nominal opening.
- Manufacturers and models
- A drawing or the leaf markings identify this gauge more reliably than a generic catalogue model.
- What we check
- Measure a blade at several places or compare a fixed opening at the contact faces. Record which interpretation applies to the asset.
- Reference equipment and accuracy
- For leaves use a calibrated micrometer with controlled contact force. For fixed gaps use calibrated end standards or a comparator of suitable uncertainty.
- When to repeat the check
- Review bent leaves, burrs and local jaw wear whenever a gap decision changes. Use that history alongside any annual recall.
Technical references CC-4643 calibration scope, page 17