Temperature, flow and detector response
Temperature indication, flow indication and detector response should have their own reference and acceptance records. Include the acquisition software revision when an exported trace is part of the evidence. That makes a later detector or controller replacement easier to follow.
A temperature value on a chromatogram may be the method setpoint written into the file, while a service worksheet records a separate observed temperature. If the job concerns agreement between the two, identify the channel and the field that appears on the printout. A method name alone does not distinguish those readings.
Gas chromatograph modules and method conditions
RTS measures the requested hardware quantities and evaluates analytical performance under the instrument protocol. Temperature, detector response and separation performance are assessed as separate characteristics, using the required reference materials and operating conditions. The record identifies which modules and methods were used for each comparison.
Specifications and service checks
A migrated analytical method may retain the original oven programme while the new instrument uses another inlet or detector. For a temperature, flow or timing enquiry, identify the actual channel used on the new unit and the part of the programme where the discrepancy occurs.
- Range and configuration
- Agilent specifies an oven operating range from 8 °C above ambient to 425 °C for its standard 8860 configuration. Flow and detector measurements have their own ranges and units.
- Manufacturers and models
- Agilent’s 8860 is a modular gas chromatograph with configurable inlets and detectors.
- What we check
- For an instrument check, identify the oven temperature, flow or timing channel to be compared. Record the column position or gas connection associated with the result.
- Reference equipment and accuracy
- Use a calibrated temperature probe and a flow reference appropriate for the gas and flow range. Specify whether reported flow is actual or converted to reference conditions.
- When to repeat the check
- Review affected results after flow-controller, oven sensor or detector work. Keep the instrument calibration history separate from the analytical method’s check results.
Technical references Agilent 8860 data sheet