On this page
Matching the hardness value to the block
Two blocks can both be marked HRC yet have quite different assigned values. Send the value, scale, serial number and certificate of the block your operator uses. A clear photograph of the working face helps establish whether there is space for further impressions. If the concern is that a particular tester disagrees with the block, identify the tester as well, so the right item receives the result.
Hardness reference blocks
Copy the complete marking and certificate identifier, including the force designation where it forms part of the scale. State whether the request concerns verification of the testing machine using that block, assessment of the block itself or replacement of an exhausted reference. These are separate jobs even when the same block is involved.
Photograph the test surface and provide the existing certificate before arranging dispatch. Used indentations reduce the area available for further readings, and the permitted spacing and usable surface come from the applicable procedure. Do not resurface or remark the test face to prepare it for calibration. Tell RTS which values, uncertainty and traceability information your quality system needs; the supported scale and reference-block arrangements must be reviewed before work is accepted.
Specifications and service checks
A bright patch between old impressions may look usable, yet the new impression must still have adequate clearance from neighbours and the block edge for that test. Keep a record of used areas before relying on a heavily tested block.
- Range and configuration
- A reference block carries an assigned hardness value for a particular scale and load. Record the full designation, such as HRC or HBW with ball size and load, and the certified usable face.
- Manufacturers and models
- Buehler offers Wilson Rockwell and Brinell reference blocks in scale-specific sets.
- What we check
- For routine verification, make properly spaced impressions on the assigned face and compare their mean with the certified value. Assignment of a new reference value requires the applicable reference-block procedure.
- Reference equipment and accuracy
- Use the block’s certification for its mean value, variation and uncertainty. A block for one Rockwell scale or Brinell combination is not a substitute for another.
- When to repeat the check
- Check the remaining usable area before each test series. Corrosion, grinding, marks on the reverse face or excessive impressions can invalidate its use as a reference.
Technical references Wilson Rockwell Test Blocks - Buehler · Wilson Brinell Test Blocks - Buehler