Frequency and drive level for inductance
Provide the working inductance points, test frequency, drive level and the selected series or parallel model. Include the fixture and compensation setting. If the bench uses both inductance and quality factor to accept a part, identify both requested readings.
A maintenance bench might check one coil at 100 Hz and another at 1 kHz. When the same coil is measured on both settings, a change in reported inductance is not automatically a meter error. Compare readings at the frequency and signal level of the assigned reference value.
Specifications and service checks
A small coil on an iron core may give a different apparent inductance at different signal levels. Include the measurement amplitude as well as frequency when that coil is used as a working reference, particularly after changing meter models.
- Range and configuration
- Specify the nominal inductance in µH, mH or H together with test frequency and level. A magnetic core may respond differently as the excitation level changes.
- Manufacturers and models
- Keysight E4980A measures impedance from 20 Hz to 2 MHz. The relevant settings include frequency, test level, fixture and series or parallel equivalent circuit.
- What we check
- Compare assigned resistance, capacitance or inductance values at the selected frequency and level. For automatic component identification, test recognition separately from value accuracy.
- Reference equipment and accuracy
- Use calibrated impedance standards characterised at the chosen frequency, with appropriate open and short corrections for the fixture. Include loss or quality factor where specified.
- When to repeat the check
- Verify with a stable check component after changing the fixture or leads and investigate shifts before changing the adjustment.
Technical references Benchtop LCR Meters: E4980A