Flow and pressure on each analyser channel
List the channels, flow settings and any pressure function used in the department. Record the units, observation time and the analyser’s displayed or exported result. This information distinguishes the calibration of the analyser from performance testing of a pump connected to it.
Infusion device analyzer as a workshop reference
Include the inlet arrangement and the manufacturer’s specified test conditions with the requested points. Calibration of this reference instrument is separate from checking an infusion pump. RTS calibrates the analyzer’s selected flow, volume, pressure and timing functions under the specified conditions.
Specifications and service checks
On a multi-channel analyser, a channel number belongs with every flow or pressure result. If only one inlet gives unusual readings, request checks of that inlet and identify its tubing rather than assigning the observation to the entire analyser.
- Range and configuration
- The Multi-Flo data lists a pressure measurement range from −500 to +2,500 mmHg. Flow, delivered volume and occlusion pressure have separate ranges. A four-channel instrument can require a result for each inlet.
- Manufacturers and models
- Rigel’s Multi-Flo measures infusion device performance and is offered with different channel counts.
- What we check
- Verify each selected channel with an independent fluid delivery or gravimetric reference. Compare the pressure indication against a separate pressure standard.
- Reference equipment and accuracy
- Use a calibrated balance and timebase or an appropriate flow transfer standard, plus a traceable pressure instrument. Choose the uncertainty against the lowest reported rate.
- When to repeat the check
- Track the results by channel. Service on a pressure sensor or fluid path warrants rechecking that channel before using it as a reference.
Technical references Rigel Medical product reference