Karl Fischer titrator qualification

A Karl Fischer instrument may use a volumetric dosing system or determine water from measured charge. The instrument type determines which equipment functions and references belong on the order. Start with the model and the result units appearing on the worksheet.

Send an enquiry
Karl Fischer titrator — equipment illustration
Equipment illustration
On this page

The dosing unit and the verification record

For a volumetric instrument, identify the dosing unit and its nominal capacity. For either system, retain the reference material certificate and the method identifier used for a whole-instrument check. This distinguishes a check of one component from a verification of the reported water result.

Karl Fischer titrator configuration

RTS evaluates the specified configuration, electrodes and any oven module against the laboratory’s protocol and water-reference material. Burette delivery is measured as a volume function on a volumetric system; complete moisture-method performance is evaluated under its own reference conditions. The report identifies which arrangement and functions were assessed.

Specifications and service checks

A laboratory using a volumetric system may have changed the dosing unit without changing the host instrument ID. Identify the dosing unit and its size when asking for a delivery check. A coulometric instrument has a different measuring principle, so copy that designation if it applies.

Range and configuration
A dosing unit has a volume capacity, whereas a titrator reports a chemical result. Volumetric and coulometric Karl Fischer instruments have different measuring principles.
Manufacturers and models
Metrohm’s 915 KF Ti-Touch is a volumetric Karl Fischer system. A coulometric instrument has a different measuring principle, so identify the model before specifying reference materials or service checks.
What we check
For a volumetric system, compare delivery from the identified dosing unit and check the relevant measuring input separately. Document any chemical performance check using an assigned reference appropriate to the method.
Reference equipment and accuracy
A calibrated balance and thermometer can support dosing-volume measurements. For a water-content check, use a suitable certified water standard with stated uncertainty, validity and units.
When to repeat the check
Review the affected functions when a dosing cylinder or electrode changes. Reagent standardisation and instrument calibration can require different schedules.

Technical references Metrohm Eco Titrator manual · Metrohm 915 KF Ti-Touch manual · ISO 8655-6 gravimetric volume determination