Sampling interval and event duration
Give the channel type, input scale, sample interval and event duration used in service. If an exported file supplies the acceptance record, specify whether its values and time stamps need checking. Identify each probe or transducer that remains part of the measurement.
A spike can appear on an oscilloscope and be absent from a recorder file if it falls between stored samples. Before attributing the difference to an amplitude error, compare the time interval between stored readings with the duration of the event.
Specifications and service checks
For a recorder triggered by an alarm, the useful evidence may lie immediately before the trigger. Specify the requested pre-trigger duration and the intended sample interval. A correct input scale cannot recover a part of the event that the recorder never stored.
- Range and configuration
- Choose input ranges and time settings from the actual recording task. A short transient can require a different timing check from a slow RMS trend.
- Manufacturers and models
- HIOKI MR8870 Memory HiCorder can record voltage waveforms and RMS data. Record the mode, channel and any transducer fitted.
- What we check
- Compare the stored sample values and elapsed time against known signals; also check displayed results if operators use them separately.
- Reference equipment and accuracy
- Use calibrated voltage and timing references with a signal source of sufficient bandwidth. Include range, trigger configuration and recording interval.
- When to repeat the check
- Recheck after changing input modules or a firmware update that affects acquisition or stored scaling.
Technical references MEMORY HiCORDER MR8870