Mixed signal oscilloscope calibration

A mixed signal oscilloscope adds logic inputs to its analogue channels. A displayed digital edge depends on the logic threshold and timing relationship as well as the acquisition timebase.

Send an enquiry
Mixed signal oscilloscope — equipment illustration
Equipment illustration
On this page

Where the logic pod detects an edge

An analogue trace crosses many voltage levels on the way from low to high. The logic pod marks one transition when the signal reaches its selected threshold. When a timing comparison is the job, tell us the analogue event the software uses for the comparison, the digital threshold, the pod model and the channels involved. Gain on the analogue input can also be checked on its own.

A logic pod and an analogue probe can display the same physical edge at different positions on the screen. Identify threshold, channel configuration and relevant delays before treating the displayed separation as the timing difference in the circuit.

Specifications and service checks

A control engineer may use an analogue channel to see a voltage ramp and a logic channel to mark when it crosses a threshold. Include the selected logic threshold with the observed delay. Changing it can shift the apparent digital transition on a slow edge.

Range and configuration
Specify analogue bandwidth and voltage ranges as well as digital threshold voltages and channel timing. Digital channel count alone says little about timing accuracy.
Manufacturers and models
Tektronix 2 Series MSO instruments combine analogue acquisition with optional digital functions. Record which digital probe and options are installed.
What we check
Compare analogue channel gain and timing, then apply known logic levels and transitions to verify selected digital thresholds and relative timing.
Reference equipment and accuracy
Use an oscilloscope calibrator or characterised signal sources and logic-level references. Include probe and cable delay in channel-alignment results.
When to repeat the check
Review after input-module or logic-probe replacement and retain a reference trace for recurring timing checks.