Probe gain at the oscilloscope input
Identify the probe type, gain or attenuation, scope input and any separate power supply. For a tip measurement, include the normal tip and lead configuration. If several probes share one scope, specify each serial number and intended channel.
A technician checking a clock through a switchable probe may see its expected frequency even if the selected attenuation is wrong. Look at the probe switch and the channel setting together whenever the timing looks correct but the displayed amplitude is too high or low.
Specifications and service checks
On a densely populated board, a long probe ground lead can add ringing to a fast transition. If the trace changes with the connection arrangement, identify the tip, attachment and return lead used. A probe attenuation check and a question about ringing call for different information.
- Range and configuration
- A 10X probe ideally delivers one-tenth of the measured voltage to the scope. Compensation range and source impedance affect response as frequency increases.
- Manufacturers and models
- Tektronix TPP0200 is a 10X passive probe with a nominal 200 MHz bandwidth when paired with a suitable input.
- What we check
- Verify attenuation at low frequency, compensate for the actual instrument input and compare frequency response to the required bandwidth.
- Reference equipment and accuracy
- Use a calibrated amplitude and frequency source and suitable reference scope. Include the probe tip, adapter and input termination in the stated setup.
- When to repeat the check
- Verify compensation when moving the probe to another input and recheck attenuation after tip or cable repairs.
Technical references TPP0100 & TPP0200 Series 100 and 200 MHz 10X Passive Probes Instructions