Passive probe calibration

A passive voltage probe relies on its divider and the input characteristics of the connected oscilloscope. Compensation and the selected attenuation position influence how a square wave appears at the display.

Send an enquiry
Passive probe — equipment illustration
Equipment illustration
On this page

Attenuation and the scope input resistance

Give the attenuation settings in use and the scope input resistance and capacitance specified for the probe. If the probe has a switchable tip or different ground accessories, identify the configuration to be checked. Note whether results are for attenuation, frequency response or both.

A probe compensated on one input can show overshoot or rounded corners when connected to a channel with different input capacitance. Keep the connection used in the comparison with the result if the probe will later be paired permanently with a particular scope.

Specifications and service checks

A switchable probe in its 1× position usually presents a different capacitive load and bandwidth from its 10× position. A circuit edge can look different even when the scope scale factor is entered correctly. Note the probe switch setting for any rise-time complaint.

Range and configuration
A 10X probe ideally delivers one-tenth of the measured voltage to the scope. Compensation range and source impedance affect response as frequency increases.
Manufacturers and models
Tektronix TPP0200 is a 10X passive probe with a nominal 200 MHz bandwidth when paired with a suitable input.
What we check
Compensate the probe for its intended scope input, then verify attenuation and pulse response. Record the cable, tip and ground accessory where high-frequency response matters.
Reference equipment and accuracy
Use a calibrated amplitude and frequency source and suitable reference scope. Include the probe tip, adapter and input termination in the stated setup.
When to repeat the check
Verify compensation when moving the probe to another input and recheck attenuation after tip or cable repairs.

Technical references TPP0100 & TPP0200 Series 100 and 200 MHz 10X Passive Probes Instructions