Attenuation and the scope input resistance
Give the attenuation settings in use and the scope input resistance and capacitance specified for the probe. If the probe has a switchable tip or different ground accessories, identify the configuration to be checked. Note whether results are for attenuation, frequency response or both.
A probe compensated on one input can show overshoot or rounded corners when connected to a channel with different input capacitance. Keep the connection used in the comparison with the result if the probe will later be paired permanently with a particular scope.
Specifications and service checks
A switchable probe in its 1× position usually presents a different capacitive load and bandwidth from its 10× position. A circuit edge can look different even when the scope scale factor is entered correctly. Note the probe switch setting for any rise-time complaint.
- Range and configuration
- A 10X probe ideally delivers one-tenth of the measured voltage to the scope. Compensation range and source impedance affect response as frequency increases.
- Manufacturers and models
- Tektronix TPP0200 is a 10X passive probe with a nominal 200 MHz bandwidth when paired with a suitable input.
- What we check
- Compensate the probe for its intended scope input, then verify attenuation and pulse response. Record the cable, tip and ground accessory where high-frequency response matters.
- Reference equipment and accuracy
- Use a calibrated amplitude and frequency source and suitable reference scope. Include the probe tip, adapter and input termination in the stated setup.
- When to repeat the check
- Verify compensation when moving the probe to another input and recheck attenuation after tip or cable repairs.
Technical references TPP0100 & TPP0200 Series 100 and 200 MHz 10X Passive Probes Instructions