Channel pairs and power factor
Give the wiring mode, channel pairs, operating frequency and power factors used on the bench. Identify any current transducers and their scaling. For a motor or inverter test, specify which side of the equipment is being measured and which quantities your report uses.
For a single sinusoidal phase at 230 V, 1 A and a power factor of 0.5, the expected active power is 115 W. The corresponding apparent power is 230 VA. Labelling the displayed field prevents the two from being compared as though they were the same result.
Specifications and service checks
A plant acceptance test may compare total three-phase power with the sum of three channels. Save the individual signs as well as the total. A reversed channel can partly cancel the others and make an otherwise significant wiring discrepancy easy to overlook.
- Range and configuration
- For sinusoidal single-phase signals, 230 V at 5 A and unity power factor is 1,150 W. At power factor 0.5 it is 575 W if voltage and current stay constant.
- Manufacturers and models
- Yokogawa's WT310E is a digital power meter. Use its input and wiring specifications to identify which quantities and channels are fitted.
- What we check
- Compare voltage, current and indicated active power at working loads and power factors. For a multichannel unit record the element numbers and phase configuration.
- Reference equipment and accuracy
- Use a calibrated power source or independent reference power meter. At low power factor, phase uncertainty can contribute more than the voltage error.
- When to repeat the check
- Review results by element. Recheck after changing an external shunt or transformer scaling.
Technical references Yokogawa WT300E series specifications