Load points and transformer ratios
List the channels and load points you need. For AC readings, include frequency and power factor. For DC, give the polarity. If the meter scales its terminal input to primary power, say which value the operator reads.
A three-phase panel can show a single total even if one current channel has its polarity reversed. Individual phase readings, where the meter provides them, are worth keeping alongside that total. One channel reading with the opposite sign points to a different issue from an error on all three.
Specifications and service checks
A machine drawing 5 kW for two hours consumes 10 kWh if its power stays constant. Identify whether purchasing needs the instantaneous power channel or the energy accumulated over a run. Some meters report both behind very similar screen labels.
- Range and configuration
- For sinusoidal single-phase signals, 230 V at 5 A and unity power factor is 1,150 W. At power factor 0.5 it is 575 W if voltage and current stay constant.
- Manufacturers and models
- Yokogawa WT310E is a line-frequency electrical power meter. Keysight N1914A is an RF power meter paired with a sensor. The input arrangement identifies which comparison applies.
- What we check
- For mains power, compare volts, amperes and active power at defined phase angles. For RF, compare indicated power against an RF reference at selected connector frequencies and levels.
- Reference equipment and accuracy
- Use a calibrated electrical power standard for mains work or a calibrated RF power transfer system for a sensor-based head. Report the reference plane and any transducer or sensor included.
- When to repeat the check
- Review results by element. Recheck after changing an external shunt or transformer scaling.
Technical references Yokogawa WT300E series specifications · N1914A EPM Series Dual-Channel Power Meter [Discontinued]