PRT sensor calibration

A platinum resistance thermometer may be delivered as a sensing element, a sheathed probe or a probe with a matched display. Identify the arrangement and the nominal resistance on the label. Reporting a resistance value and reporting temperature on a paired display are different requests.

Send an enquiry
PRT sensor — equipment illustration
Equipment illustration
On this page

Sensor coefficients and the readout

If the sensor is used with stored coefficients, provide the coefficient set or readout identity. A probe can behave consistently while two readouts show different temperatures because they are using different conversion settings.

If a technician transfers the probe to another readout, the new display may use a generic resistance curve instead of the previous probe-specific coefficients. A sensor certificate giving resistance values and a record of the active coefficients let the laboratory investigate that change without losing the sensor’s history.

Specifications and service checks

A laboratory may enter individual probe coefficients in one reader while a second reader uses a stock platinum curve. If the same probe gives different temperatures, save the configuration from each reader. The comparison can then separate probe resistance from the conversion that produced each display.

Range and configuration
Pt100 and Pt1000 identify nominal resistances of 100 Ω and 1,000 Ω respectively at 0 °C. The sheath, element construction and stated IEC 60751 tolerance class determine the useful temperature band.
Manufacturers and models
WIKA documents Pt100 and Pt1000 resistance assemblies in its TR10 family. Confirm the element, tolerance class and wiring from the instrument record.
What we check
Compare the sensor with a reference thermometer and report resistance or temperature error at the required points. Include its wire configuration and any transmitter fitted in the head.
Reference equipment and accuracy
A stable temperature source, traceable reference thermometer and resistance readout establish a comparison. Quantify self-heating and lead effects if they are material to the requested PRT tolerance.
When to repeat the check
Track resistance at a repeatable check point. Mechanical shock, high-temperature service and a change in lead configuration are reasons to recheck sooner.

Technical references WIKA TR10 and TC10 operating instructions · BIPM guide to industrial platinum resistance thermometers · Endress+Hauser temperature measurement