On this page
Sensor coefficients and the readout
If the sensor is used with stored coefficients, provide the coefficient set or readout identity. A probe can behave consistently while two readouts show different temperatures because they are using different conversion settings.
If a technician transfers the probe to another readout, the new display may use a generic resistance curve instead of the previous probe-specific coefficients. A sensor certificate giving resistance values and a record of the active coefficients let the laboratory investigate that change without losing the sensor’s history.
Specifications and service checks
A laboratory may enter individual probe coefficients in one reader while a second reader uses a stock platinum curve. If the same probe gives different temperatures, save the configuration from each reader. The comparison can then separate probe resistance from the conversion that produced each display.
- Range and configuration
- Pt100 and Pt1000 identify nominal resistances of 100 Ω and 1,000 Ω respectively at 0 °C. The sheath, element construction and stated IEC 60751 tolerance class determine the useful temperature band.
- Manufacturers and models
- WIKA documents Pt100 and Pt1000 resistance assemblies in its TR10 family. Confirm the element, tolerance class and wiring from the instrument record.
- What we check
- Compare the sensor with a reference thermometer and report resistance or temperature error at the required points. Include its wire configuration and any transmitter fitted in the head.
- Reference equipment and accuracy
- A stable temperature source, traceable reference thermometer and resistance readout establish a comparison. Quantify self-heating and lead effects if they are material to the requested PRT tolerance.
- When to repeat the check
- Track resistance at a repeatable check point. Mechanical shock, high-temperature service and a change in lead configuration are reasons to recheck sooner.
Technical references WIKA TR10 and TC10 operating instructions · BIPM guide to industrial platinum resistance thermometers · Endress+Hauser temperature measurement