Pt1000 sensor calibration

A Pt1000 has a nominal resistance of 1000 ohms at zero degrees Celsius. It needs a readout configured for that sensor type. A replacement display intended for a Pt100 can therefore cause a problem even though both probes are platinum resistance devices.

Send an enquiry
Pt1000 sensor — equipment illustration
Equipment illustration
On this page

Probe resistance or the paired display reading

Record the wiring arrangement and identify the display or transmitter normally connected. A probe-only result in ohms can be reused with another suitable reader, while a system result belongs to the particular sensor and readout tested together.

When replacing a building sensor, the transmitter configuration is as important as the connector. A spare marked Pt1000 should be recorded with the input type selected on the transmitter. Otherwise a wiring or configuration fault could be treated as an error in the new sensor.

Specifications and service checks

A Pt1000 has a nominal resistance of 1,000 Ω at 0 °C. A universal input left on its Pt100 setting expects 100 Ω at that temperature. After a probe change, compare the input selection with the nameplate before drawing conclusions from an implausible display.

Range and configuration
A Pt1000 has a nominal 1,000 Ω resistance at 0 °C. Identify the curve, connection and tolerance class of the installed element. Assembly construction determines the usable temperature range.
Manufacturers and models
WIKA documents Pt100 and Pt1000 resistance assemblies in its TR10 family. Confirm the element, tolerance class and wiring from the instrument record.
What we check
Compare the sensor with a reference thermometer and report resistance or temperature error at the required points. Include its wire configuration and any transmitter fitted in the head.
Reference equipment and accuracy
Use a calibrated platinum reference, controlled bath or block and a traceable resistance readout. A Class A or B label describes a tolerance envelope; a reference needs known uncertainty and corrections.
When to repeat the check
Track resistance at a repeatable check point. Mechanical shock, high-temperature service and a change in lead configuration are reasons to recheck sooner.

Technical references WIKA TR10 and TC10 operating instructions · BIPM guide to industrial platinum resistance thermometers · Endress+Hauser temperature measurement