Weighing in air and in liquid
For a check of the reported relative density, send the suspension or basket used for the immersed weighing, the reference item and the liquid-temperature convention in your procedure. For example, changing the liquid setting in the display may change the density result even when the two underlying mass readings agree. Keeping those readings with the calculated result makes that difference visible.
Specific gravity balance and immersion accessories
A mass calibration addresses the weighing function; the calculated density also depends on the defined comparison and conditions. Supply the laboratory method and temperature requirement; RTS measures the requested mass quantities and checks the specified density calculation separately.
Specifications and service checks
A balance can agree with a check weight in air while its submerged weighing arrangement introduces a separate difference. If the request concerns the density function, identify the suspension, sinker or basket and the liquid temperature as well as the balance. The two weighings need an identifiable setup.
- Range and configuration
- The weighing capacity and readability limit which samples can be used, but the density or specific-gravity uncertainty also depends on immersion and the liquid temperature.
- Manufacturers and models
- OHAUS lists solids density kit 80253384 as an accessory for its Explorer precision balances. Identify the actual balance and the kit separately if the suspended weighing arrangement forms part of the check.
- What we check
- Check the balance indication with certified weights. For hydrostatic weighing, assess the complete density arrangement with an appropriate reference and the reported temperature.
- Reference equipment and accuracy
- Choose weights with uncertainty suitable for the balance and a traceable thermometer for the liquid. Reference-liquid density and buoyancy contributions belong in a density result.
- When to repeat the check
- Review after replacing the suspension or immersion kit. A passing mass check alone may miss a change in the density setup.
Technical references OHAUS Explorer precision balances · OIML guide to density measurement