On this page
Reading a load as it comes off the scale
A pointer may show one value as a load is added and a slightly different value at the same load on the way down. If the scale is used while a suspended container empties, the descending readings matter to your operator. Tell us the scale graduation and which loads you need covered. A note on where the unloaded pointer ends up also helps us interpret the small-load readings.
Spring balance: mass or force indication
Copy the complete scale and describe how the load is applied. Give the points used and whether readings during loading and unloading are needed. RTS calibrates the spring balance in the specified mass or force units and operating orientation.
Specifications and service checks
A hanging scale graduated in newtons displays force. A 1 kg mass exerts about 9.81 N under standard gravity. Clarify the unit if the spring balance is used on one job for pulling force and another for weighing.
- Range and configuration
- A spring scale can be graduated in newtons or in mass units. Pesola’s Micro Line includes a 100 g model with 1 g divisions. Read the unit on the actual spring before specifying loads.
- Manufacturers and models
- Pesola 20100 is a mechanical Micro Line spring balance rated to 100 g. Give the scale division and any removable pan or hook used to apply the load.
- What we check
- Apply traceable loads in the normal hanging position and compare rising and falling readings. On a spring scale record the unloaded pointer on return.
- Reference equipment and accuracy
- Use calibrated masses for mass indication or a suitable force reference for force. The local acceleration of gravity matters when masses are used to realise force.
- When to repeat the check
- Check the unloaded pointer and a known mass between calibrations. A permanently stretched spring, distorted hook or zero that changes after loading warrants an earlier check.
Technical references Pesola Micro Line spring balances · BIPM definitions of mass and weight