The sensing element and its bridge circuit
Describe whether the item is a bonded gauge, a completed force sensor or an indicator with a bridge input. Give the gauge factor or rated output where available. For a result in force units, identify the assembled load path and the points at which force is applied.
A bridge specified at 2 mV/V produces a nominal 10 mV at rated load with 5 V excitation. Record the actual excitation when interpreting a millivolt reading. Changing the supply can change raw output without changing the applied load.
Specifications and service checks
A quarter-bridge strain channel may be zeroed each time a fixture is changed. If a complaint began with a longer extension cable, identify the bridge arrangement and lead compensation. Cable resistance can enter the measurement differently depending on that arrangement.
- Range and configuration
- The resistance marked on a gauge and its gauge factor are distinct. The measured quantity in use is strain, typically expressed in microstrain.
- Manufacturers and models
- HBK LY-series linear strain gauges are supplied with different nominal resistances and grid dimensions.
- What we check
- Compare the installed bridge output against known strain or a suitable mechanical reference. An electrical shunt check only tests part of the measuring chain.
- Reference equipment and accuracy
- Use a calibrated strain reference or appropriate extensometer for an applied-strain comparison. A traceable resistor can support a separate bridge check.
- When to repeat the check
- Review after rebonding the gauge, changing lead compensation or replacing the amplifier.
Technical references HBK LY-series strain gauges