On this page
Pen readings and paper travel
Include a piece of the ruled paper or its part number, the chart speed and the input range selected for each pen. When comparing old and new charts, check the printed paper scale before interpreting a change in the height of a trace.
At a selected paper speed of 10 mm per minute, a feature spanning 30 mm represents three minutes if the paper drive performs as set. If elapsed time is part of your acceptance decision, request a chart travel check alongside the input-to-pen reading.
Specifications and service checks
A roll advance of 20 mm per hour puts two events 30 minutes apart about 10 mm apart on the paper. If the spacing looks wrong, identify the selected chart speed and paper type. This tests the time axis independently of where the pen sits on the value scale.
- Range and configuration
- The pen’s travel represents the configured input span. A chart marked in degrees may be driven by a thermocouple, a resistance input or a transmitter.
- Manufacturers and models
- Yokogawa’s µR10000 series includes pen recorders with selectable chart speed.
- What we check
- Apply a known input and check the trace against the correct chart divisions. Mark start and finish times on the strip if paper advance or elapsed time forms part of the order.
- Reference equipment and accuracy
- Use a calibrated electrical source matched to the input and an independent time reference. Paper scale, pen width and alignment limit the reading.
- When to repeat the check
- Renew the comparison after a range change or pen mechanism repair. Retain a marked chart to compare with the next visit.
Technical references Yokogawa µR10000 recorders