On this page
Matching the foil or shim to its value
Mark each piece with its nominal value or retain its individually labelled envelope. For a coated standard, identify the substrate and usable area. When only part of a patch is used by the probe, note that position rather than treating its entire area as identical.
Thickness standard and calibration foil
List its assigned value, material, identifier and the gauge with which it is used. CC-4643 includes a standard-foil entry; identify the foil or other thickness reference by its construction and assigned value. RTS measures the specified reference thicknesses and records any damaged or creased areas that affect use.
Specifications and service checks
A loose thickness foil placed on a curved substrate can lift away from the surface. If it serves as a coating-gauge check, keep the foil thickness and the substrate geometry together. Measuring the foil alone answers only the length question.
- Range and configuration
- For a loose foil the assigned quantity is physical thickness. A coated base can instead have an assigned coating thickness. The RTS standard-foil entry covers 0.05 to 2 mm.
- Manufacturers and models
- Elcometer supplies calibration foils for coating gauges. Give the foil’s individual value and the type of base normally used.
- What we check
- Measure the foil at specified locations with controlled contact force. For a coated artefact retain its assigned substrate, area and verification method.
- Reference equipment and accuracy
- A traceable contact comparator can assess a suitable loose foil. Use a substrate-appropriate certified standard for checking a coating measurement system.
- When to repeat the check
- Replace or recheck creased foils. Repeated measurements on the same small area can change the value of a soft reference.
Technical references Elcometer 456 Integral Coating Thickness Gauge · CC-4643 calibration scope, page 19