Adding short run periods on the register
A totaliser marked 0.1 h changes its last digit every six minutes of accumulated run time. An individual two-minute cycle may therefore finish without a visible change on the register. For a job involving short cycles, give the input contact and the usual on-time. If the register is expected to remember the total through a power loss, mark that requirement as well.
A machine totaliser may start when a motor contact closes, while a shift clock includes idle time as well. Compare the register with the sum of the actual active periods. Otherwise normal machine stops may be counted as a timing error.
Specifications and service checks
A packaging line may stop and restart several times each shift. If the totalizer drops part of a minute on every restart, a single uninterrupted hour comparison could miss the complaint. Include an example of the usual on/off sequence.
- Range and configuration
- The least display increment matters. A register that advances in tenths of an hour has a 6-minute step, even if the internal timebase is much finer.
- Manufacturers and models
- Crouzet makes electromechanical elapsed-time counters. Record whether the device counts supply-on time or time gated by a separate input.
- What we check
- Apply independently timed enable periods, including a pause, and compare the accumulated register change. Record what the totaliser does through a power interruption if that affects its use.
- Reference equipment and accuracy
- Use a traceable timebase or reference timer with an unambiguous start and stop. The display step sets a lower limit to useful test duration.
- When to repeat the check
- Review drift over a long interval and examine the run input after wiring changes.
Technical references Crouzet elapsed-time counters