On this page
Identifying pressure, load and displacement channels
On one frame, a 1 MPa pressure channel and a 1 kN load channel might each have a display marked ‘1.000’. List them by function, sensor tag and unit. If a pressure transducer moves between cells, the old certificate can still identify the transducer, while the new report needs to show the input it used. Include any axial displacement and volume channels in the same channel list.
Triaxial test apparatus
State the test mode and which channels enter the calculation. Include the cell, controller and software identifiers. The proposed assessment must define the individual measurement chains; verifying pressure alone does not establish the complete system's performance for a soil test.
Specifications and service checks
Cell pressure and axial load can be logged in neighbouring columns. Keep the channel map and units from the data logger with the job, especially after a transducer swap or software reconfiguration.
- Range and configuration
- For example, 25 kN and 50 kN load cells appear in the system options. Cell pressure, pore pressure and axial displacement have separate ranges.
- Manufacturers and models
- Matest TriaxLab systems can be supplied with different load cells and displacement transducers.
- What we check
- Identify the cell and channels. Verify axial load and travel, then compare each specified pressure channel with a pressure reference and retain its identity.
- Reference equipment and accuracy
- Use force, displacement and pressure standards with uncertainty appropriate to their individual tolerances. One frame-capacity figure cannot qualify every channel.
- When to repeat the check
- Review the affected measurements after changing a cell, transducer or readout. Keep separate histories for interchangeable sensors.
Technical references Matest product reference