Transformer ratios and the sign of reactive power
Identify the network arrangement, transformer ratios, frequency and sign convention. Give representative inductive and capacitive points if both appear in service. A result stated in secondary var should be clearly distinguished from one scaled into primary kvar on a switchboard.
A switchboard may use a sign to distinguish reactive power supplied from reactive power absorbed. An apparently equal magnitude on the meter and a portable reference can still represent opposite directions. Record the direction convention on both instruments when reviewing the comparison.
Specifications and service checks
A capacitor bank controller can cross from inductive to capacitive operation as stages switch in. If the plant responds to both directions, include one example of each and the sign convention. A magnitude-only comparison misses a reversed indication.
- Range and configuration
- For sinusoidal waveforms, power factor relates to the cosine of phase angle, while reactive power is reported in var or kvar. Quote the wiring and sign convention.
- Manufacturers and models
- Rishabh's Rish Delta power-meter family illustrates electrical power derived from voltage and current. Identify the installed function as power factor or reactive power.
- What we check
- Apply known reactive power at leading and lagging phase displacement, and verify which direction gives positive var on the chosen wiring. A unity-power-factor condition provides a useful near-zero check.
- Reference equipment and accuracy
- Use a calibrated multiphase power source or reference power meter. For readings close to zero reactive power, phase uncertainty is especially important.
- When to repeat the check
- Reassess after CT reconnection, phase sequence changes or a new scaling factor.
Technical references Rishabh Rish Delta power-meter manual