Port configuration and S-parameters
Give the port configuration, connector interfaces, frequency sweep and parameters used by the laboratory, such as forward transmission or input reflection. Include the calibration kit or module and any fixtures when the request covers the measurement system beyond the instrument ports.
If a cable is added after the reference plane has been established, its electrical length appears in the phase result. A comparison made directly at the ports cannot be treated as equivalent to one made at the end of that cable without accounting for the change.
Specifications and service checks
A phase result may change after a test cable is flexed while its amplitude remains almost unchanged. Identify the cable and fixture that stay with the port during use. That helps separate stability of the accessory path from receiver amplitude performance.
- Range and configuration
- Specify the S-parameters, frequency span, port impedance and required dynamic range. S11 describes reflection at port 1, while S21 describes forward transmission.
- Manufacturers and models
- Keysight's E5061B ENA network analyser is available in different test-set configurations. Quote the exact option and connectors.
- What we check
- After port error correction, verify both the magnitude and phase of relevant S-parameters against independent characterised devices. Cable movement after correction can invalidate the established plane.
- Reference equipment and accuracy
- Use compatible calibrated standards and an independent verification set. At high loss, noise floor and residual leakage can dominate uncertainty.
- When to repeat the check
- Verify performance after replacing a test cable or disturbing a connector and review formal calibration against the drift record.
Technical references E5061B ENA Series Vector Network Analyzer