Frequency, polarity and voltage convention
State the operating levels and polarity for DC, or amplitude convention and frequency for AC. Include the normal load or input resistance where it can affect the result. For a bipolar calibrator, list both positive and negative values used in the acceptance procedure.
A controller input set for a bipolar range should be checked on both sides of zero if it uses both in service. Agreement at a positive setting alone does not characterise negative output. Keep any offset around zero visible instead of folding it into an unsigned error.
Specifications and service checks
For a sensor input that normally operates between 1 and 5 V, the useful midpoint is 3 V. A calibrator’s 0–10 V range can accommodate the signal, but the operator’s three working voltages convey the intended application much better than the range name.
- Range and configuration
- For a voltage-only calibrator list the smallest and largest outputs used, DC polarity and AC frequencies. At millivolt levels include the output and lead configuration used for zero.
- Manufacturers and models
- Fluke 5560A is a multiproduct calibrator with several electrical output functions. A voltage-only asset can be assessed against a subset of these functions.
- What we check
- Measure assigned outputs independently at the requested points and loads. Keep the results for voltage, current, resistance or frequency distinguishable by function.
- Reference equipment and accuracy
- Use reference meters, characterised shunts or standards suited to each output. Evaluate the reference uncertainty against the individual range specification.
- When to repeat the check
- Monitor key outputs with check standards between calibrations. Review the interval against drift of the functions that set the tightest tolerance.
Technical references Fluke 5560A calibrator family specifications