Wattage on the selected current and voltage ranges
List the AC or DC function, rated input ranges, working power and any external current sensor. Include frequency and the leading or lagging condition for AC tests. Identify the scale multiplier where an analogue wattmeter uses switchable current and voltage ranges.
An analogue wattmeter can reach the current-coil limit even when the pointer reads only a fraction of full scale on a low-power-factor load. If the dial has several multipliers, include the selected voltage and current ranges alongside the observed division.
Specifications and service checks
For a 100 V, 2 A single-phase sinusoidal load at unity power factor, the expected active power is 200 W. At a power factor of 0.5 it is 100 W. Equal voltage and current readings alone therefore do not establish the watt indication.
- Range and configuration
- Record watts and the separate voltage and current input ratings. The same wattage can occur at different input levels and power factors, giving different phase-related errors.
- Manufacturers and models
- Yokogawa's WT310E is a digital power meter. Use its input and wiring specifications to identify which quantities and channels are fitted.
- What we check
- Compare voltage, current and indicated active power at working loads and power factors. For a multichannel unit record the element numbers and phase configuration.
- Reference equipment and accuracy
- Use a calibrated power source or independent reference power meter. At low power factor, phase uncertainty can contribute more than the voltage error.
- When to repeat the check
- Review results by element. Recheck after changing an external shunt or transformer scaling.
Technical references Yokogawa WT300E series specifications