Weighing indicator calibration

A weighing indicator converts a load-cell bridge signal into a displayed mass. Stored tare, multiple ranges and the number of divisions can all affect the reading presented to the operator.

Send an enquiry
Weighing indicator — equipment illustration
Equipment illustration
On this page

Gross mass, net mass and the indicator input

If a vessel is tared before filling, the operator usually reads net mass. A service engineer may be looking at gross mass on the same display. Tell us which value is used and give the division and range. For an indicator-only check we need the bridge input corresponding to full scale. If the question is whether the platform weighs correctly, the loading points, installed cells and display must be considered together.

Weighing indicators and the scale assembly

Record the input sensitivity, configured capacity, division and load-cell arrangement. State which assembly is being supplied and which quantity should appear on the report. RTS records whether the result is for the electrical indicator or for the complete weighing assembly under applied mass.

Specifications and service checks

On a batching skid, a display step of 0.5 kg can mask a small change in load-cell signal. Ask whether the comparison is to cover the electrical input, the displayed values or the full weighing assembly under load. Those choices determine what the check can resolve.

Range and configuration
A bridge indicator receives a small signal, commonly expressed in mV/V, and displays the configured engineering span. The digital capacity alone does not identify the input range.
Manufacturers and models
Rice Lake 920i is a weighing indicator/controller for load-cell systems. Identify the particular load input and displayed unit.
What we check
A bridge simulator checks the indicator electronics. Where the result must include the load cell and structure, specify an applied force or applied mass comparison of the full system.
Reference equipment and accuracy
A calibrated bridge simulator can verify mV/V response at the indicator terminals. A complete scale comparison also needs appropriate calibrated masses and includes the load cell and structure.
When to repeat the check
Review after changing excitation, scaling, the load cell or the mechanical mounting. Keep results for electronics-only checks identifiable.

Technical references 920i Programmable Indicator/Controller Technical Manual