Two altimeters can show different elevations at the same point if their barometric reference settings differ. A pressure-based altimeter needs that setting recorded alongside the applied pressure and indicated altitude.

Send an enquiry
Altimeter — equipment illustration
Equipment illustration
On this page

Benchmarks and the pressure reference setting

If the field crew noted an error at a survey benchmark, copy the benchmark elevation, displayed altitude and pressure reference setting from the field book. With a barometric altimeter, changing the reference can move the indicated altitude while the instrument stays on the same bench. A satellite-derived height involves a different comparison. Give us the model and mode before booking the check.

Altimeter measurement principle

A barometric model needs its pressure reference, units and configured reference conditions identified. State whether pressure indication or the derived altitude display requires assessment. RTS calibrates the specified altitude-related indication with the operating principle and reference conditions recorded; any aviation-specific acceptance requirement is identified separately.

Specifications and service checks

A survey crew can see its indicated height change at a fixed benchmark as air pressure changes. Record the pressure reference setting and time alongside the benchmark elevation. This makes the return visit a meaningful comparison.

Range and configuration
Suunto lists an altitude display from −500 to 9,000 m in 1 m increments. Its altitude indication is derived from pressure and the reference setting. Give the elevation band used and the pressure reference convention.
Manufacturers and models
Suunto Core is an example of a portable barometric altimeter.
What we check
Compare the response to traceable pressure points and document the reference setting used to convert pressure to height. A single known elevation checks only one point.
Reference equipment and accuracy
Use a pressure standard with uncertainty sufficient for the claimed height tolerance. Record the conversion convention and any rounding of the display.
When to repeat the check
Review after sensor work or an unexplained discrepancy against a known elevation. Field zero checks complement periodic calibration.