Clinometer calibration

A clinometer reads the slope of the surface supporting its base. A fixture set to a positive angle and the same fixture inclined the other way use opposite sides of zero. A check at level alone will miss errors that appear farther up either scale.

Send an enquiry
Clinometer — equipment illustration
Equipment illustration
On this page

Reading a slope in degrees or millimetres per metre

On a machine bed, a rise of 1 mm over a horizontal metre is about 0.057 degrees. It is easy to misread that figure if one shift records millimetres per metre and another records degrees. Give us the unit used on your inspection sheet, the normal orientation of the base and the positive and negative slopes that matter for that bed.

Clinometer axes and reference faces

Identify the axis or axes requiring results and any separate sensor and display. State whether the instrument reads an angle directly or a slope in another unit. RTS calibrates the selected axes and points against the stated reference faces and mounting arrangement.

Specifications and service checks

A ramp can rise in the direction of travel and also fall across its width. Mark which edge of the instrument follows the direction being measured. Rotating the base through a quarter turn changes the slope under inspection.

Range and configuration
WYLER gives ±60° as the measuring range for Clinotronic PLUS. For a fixed angle block, quote the angle between its working faces instead.
Manufacturers and models
WYLER Clinotronic PLUS is a portable electronic instrument. Its range and readout apply to this design.
What we check
Take readings of the same tilt in forward and reversed orientations. If the clinometer is used on a cylindrical surface, identify the V-base and the measurement direction.
Reference equipment and accuracy
A calibrated angular table, sine setup or angle blocks establish the reference. Specify the base used, its orientation and the required uncertainty in degrees or arcminutes.
When to repeat the check
A fall or a marked change on reversal calls for a fresh check. Periodically check zero in the position used on the machine.

Technical references WYLER Release 2.0 · WYLER Angle gauge reference