On this page
Recognising a component and measuring its value
A 100 nF capacitor might be recognised correctly on every attempt. The question for an inspection bench is how close the indicated capacitance is at its working frequency and test level. Send the component value or semiconductor parameter your operators accept or reject, and the fixture used to make contact. If the automatic component-identification function also matters, we can record that check in its own right.
Component testers with automatic results
Identify the functions and component types actually used for acceptance. Keep fixture, test level, frequency and equivalent-circuit settings with each requested result. RTS checks the selected electrical parameters at the stated test levels and frequencies. Automatic component classification and semiconductor functions have their own checks; list those results separately from resistance, capacitance and inductance measurements.
Specifications and service checks
For an incoming-inspection jig, a component tester might identify polarity correctly yet grade parts by an electrical parameter at a particular test condition. Name the grading parameter and condition. A general identification check and a quantitative part-grading comparison have different aims.
- Range and configuration
- A nominal 1 µF capacitor can give different results at 100 Hz and 1 kHz. For inductors, both frequency and test current can influence the reported inductance.
- Manufacturers and models
- Keysight E4980A is an example of an impedance instrument for passive components. Automatic identifiers and semiconductor analysers measure other properties, so record the component types the unit claims to test.
- What we check
- Verify assigned resistance, capacitance or inductance at the selected frequency and level. For a diode or transistor function, use characterised devices and the specified operating current or voltage. Verify automatic component recognition separately.
- Reference equipment and accuracy
- Use calibrated impedance standards for passive parts and characterised semiconductor references or a suitable source-measure reference for electrical device parameters.
- When to repeat the check
- Verify with a stable check component after changing the fixture or leads and investigate shifts before changing the adjustment.
Technical references Benchtop LCR Meters: E4980A