Diaphragm seal inspection and setup

A diaphragm seal transmits process pressure through a fill fluid to the gauge or transmitter. Seal elevation and capillary arrangement can affect the zero of the combination, especially on a low pressure range.

Send an enquiry
Diaphragm seal — equipment illustration
Equipment illustration
On this page

Height difference between the seal and transmitter

When a transmitter is mounted below its seal, the fill fluid contributes a pressure head. A comparison with both parts level on a bench can therefore give a different zero from the installed one. A sketch showing the installed heights, the capillary length and the model of each part is more useful than a photograph of the transmitter alone. Send the sealed assembly intact if you want the result to include it.

Diaphragm seals and attached transmitters

Keep the seal, capillary and transmitter identified as one assembly when that is how they are used. Give their relative mounting positions and the operating temperature information needed for review. RTS calibrates the filled assembly in its specified configuration, retaining the relative positions and the pressure reference with the results.

Specifications and service checks

A transmitter shipped alone after removal from a remote seal no longer represents the installed sealed system. Identify the matching seal, capillary length and orientation as installed when the question concerns a reading on a vessel.

Range and configuration
The working pressure band follows the assembled system. Ambient and process temperatures can affect the result through the fill fluid.
Manufacturers and models
WIKA 990.27 is a flanged diaphragm seal used as part of a filled pressure assembly. Note the attached gauge or transmitter and capillary length.
What we check
Compare the assembled seal and indicator against known pressures with the seal and readout at recorded elevations. Check both increasing and decreasing pressure.
Reference equipment and accuracy
Use a calibrated pressure reference and compatible means of applying pressure to the diaphragm. Account for fill-fluid head and temperature where material.
When to repeat the check
Recalibrate the assembly after refilling, changing the capillary or replacing the measuring instrument.