When the plunger lock slips, the span can change while the gauge travels from the bore to the micrometer. A useful quotation therefore names the particular heads and working sizes, along with the setting rings and the outside measuring instrument used at your bench. If only one head in a set gives trouble, mark it on the case so it can be matched to the comparison.
Telescopic gauge transfer measurements
Identify its working interval, contacts and locking arrangement, together with the micrometer or other instrument used to read it. State the defined comparison required. RTS checks the transfer measurement against a dimensional reference, recording the contact and locking arrangement.
Specifications and service checks
A two-ended transfer head used across a round hole needs its span transferred to a micrometer after withdrawal. If the micrometer reads differently when a second operator makes the transfer, compare the same locked setting before resetting in the bore.
- Range and configuration
- The nominal band on each head is the bore diameter it can reach. A telescoping gauge transfers that size to a micrometer and does not display bore size itself.
- Manufacturers and models
- Mitutoyo Series 155 is a telescoping gauge set. Quote the individual head range and the micrometer used to read it.
- What we check
- Check the transfer in a calibrated setting ring near the sizes used, including locking repeatability and the condition of the contact tips.
- Reference equipment and accuracy
- Use a certified setting ring and a calibrated outside micrometer capable of resolving the bore tolerance. Report both parts of the transfer arrangement.
- When to repeat the check
- Test withdrawal and retention after a dropped head or lock repair. Frequent transfer checks give a better indication of wear than calendar age alone.