On this page
Following a measurement through the bench
Provide the bench drawing or channel list and one example of the result that the operator signs. Mark which channels source signals, which measure them and which fixtures remain connected. This lets the job cover either named instruments or the complete working path.
The electrical test bench as a measurement chain
Provide the circuit or connection diagram, instrument list and the quantities used to accept the product. Separate calibration of those instruments from verification of the complete bench sequence. If software applies corrections, retain its revision and correction file with the request. RTS calibrates the individual instruments and verifies the defined measurement sequence, with the references and software settings recorded.
Specifications and service checks
On a multi-station bench, one controller may feed separate fixtures with their own transducers. Mark the station numbers where parts are accepted and the reading recorded for each. A common display can hide differences introduced at the individual stations.
- Range and configuration
- The bench may combine pressure, voltage, current and resistance with different spans. List each production result with its range and governing tolerance.
- Manufacturers and models
- Beamex CENTRiCAL illustrates a bench built from selected calibration modules. Identify the actual modules, switching and fixture in the operator's signal path.
- What we check
- Verify the outputs and inputs individually, then compare a representative measurement through the complete configured path, including any software scaling.
- Reference equipment and accuracy
- Use independent references for each measured quantity. Include cabling, switching and the fixture when estimating uncertainty at the workpiece connection.
- When to repeat the check
- Recheck affected paths after module replacement, fixture repair or changes to calculation software.
Technical references Beamex CENTRiCAL calibration bench for the process industries