Variac calibration

A Variac changes its output through a moving contact on an autotransformer winding. The dial tells you where the contact sits. The actual output voltage also depends on the supply voltage and the load connected.

Send an enquiry
Variac — equipment illustration
Equipment illustration
On this page

Output voltage under the specified load

State the nominal input, frequency, working output points and load condition. If the unit has its own voltmeter, ask for the output-to-dial relationship and the meter indication as separate results. Identify any external isolation arrangement independently of the variable transformer.

Two job sheets may both say “half scale” even though one technician read the dial mark and the other read a panel voltmeter. Record the knob position, input voltage and measured output under the specified load together. A fitted voltmeter can then be checked against that same output.

Variac and manually variable AC supplies

Give the supply condition, requested output points and load arrangement. A dial position is an adjustment setting unless its relationship to delivered voltage is part of an agreed measurement. List the external reference or indicator separately, and ask RTS to confirm the specific voltage checks rather than treating the whole assembly as a multifunction calibrator.

Specifications and service checks

An adjustable transformer on a test bench may have a dial in percent. A 50% dial mark is a setting, while the actual volts depend on the supply and loading. Give the supply condition and working load when a particular output voltage is important.

Range and configuration
The output span depends on the winding and supply voltage. Some units extend above nominal input. The knob's percentage scale is not an independent voltmeter.
Manufacturers and models
A Variac is a variable autotransformer. Record the manufacturer, winding connection and rated input rather than identifying one by the dial position alone.
What we check
Compare terminal voltage with the dial or displayed indication at agreed settings under documented input and load conditions. Include the return direction if repeatability matters.
Reference equipment and accuracy
Use calibrated AC voltage measurement on input and output. A stable source and specified load let the report distinguish ratio error from supply variation.
When to repeat the check
Review after brush replacement, erratic adjustment or winding repair, along with the normal electrical maintenance schedule.