Thermistor calibration

A thermistor can have a steep, non-linear resistance change with temperature. Its nominal value at a stated temperature and its conversion data identify the element. Two probes with the same connector may give quite different indications on the same electronics.

Send an enquiry
Thermistor — equipment illustration
Equipment illustration
On this page

Sensor resistance or the paired temperature reading

If you need the corrected temperature as it appears on the display, send the thermistor with that readout. For a sensor-only comparison, request resistance at the chosen temperatures and identify any table or coefficients the results will be compared against.

A replacement probe marked 10 kΩ can match the old one at its nominal temperature yet follow a different curve elsewhere. The controller may retain the curve selected for the original probe. Include the probe part number, curve or coefficients where available and the controller configuration if you need to diagnose that mismatch.

Thermistors with or without readouts

Identify the sensor curve or coefficients used, the cable arrangement and the requested temperature points. Keep the readout settings with the sensor identity where they form a pair. RTS calibrates the thermistor using its specified resistance–temperature relationship, with any paired readout included in the results.

Specifications and service checks

Two probes both marked 10 kΩ at a reference temperature can follow different resistance curves away from that point. If a replacement probe agrees near room temperature but differs elsewhere, obtain its curve designation and the curve selected in the reader.

Range and configuration
Fluke specifies 0 to +100 °C for its 5611A reference probe. A different thermistor can have a different curve and usable interval even when it has the same resistance at a single temperature.
Manufacturers and models
Fluke Hart 5610 and 5611A are examples of secondary reference thermistor probes.
What we check
Compare the sensor in a stable medium with a traceable reference and record resistance at each temperature. If it normally has a dedicated readout, identify the pairing on the report.
Reference equipment and accuracy
Use a calibrated reference thermometer and resistance meter with suitable measuring current. Self-heating and interpolation can matter more than the last display digit.
When to repeat the check
Review check-point resistance and condition of the lead before extending an interval. Reassess if the probe or a programmed coefficient set changes.