UV lamp irradiance checks

A UV lamp is a source of radiation rather than an instrument that automatically measures its own output. An output check requires an agreed measurement position, wavelength band and reference detector. A lamp-hours display supplies different information from measured irradiance.

Send an enquiry
UV lamp — equipment illustration
Equipment illustration
On this page

Measurement distance and detector position

Identify the lamp and housing, the distance or fixture used for measurement and any window in the optical path. Keep those details with the output reading. After a lamp replacement you can repeat the comparison at the same distance and detector position.

If your old report gives an output reading with the cover in place, a later reading taken with the cover removed answers a different question. Include the cover or window in the configuration description and identify the detector’s wavelength band. This is particularly useful when the lamp itself has been replaced but the fixture has stayed in service.

UV lamp and irradiance measurements

Identify the lamp type, required spectral response, measurement position and the purpose of the engineering check. Lamp operating time alone is not an irradiance result. Ask for the suitable specialist method to be reviewed; a lamp check does not by itself establish a validated decontamination process.

Specifications and service checks

A replacement lamp may fit the same holder while the irradiance at a work surface depends on position. Mark the distance and receiving plane, and copy the wavelength band from the application. A power marking on the lamp cannot supply the irradiance at that plane.

Range and configuration
Analytik Jena lists a 302 nm transilluminator option for its UVP ChemStudio. Wavelength identifies the band. For an irradiance result, specify the measurement plane and units such as mW/cm².
Manufacturers and models
Analytik Jena’s UVP ChemStudio family includes UV transilluminator options with different light-source arrangements.
What we check
Compare irradiance at a defined measurement plane with a calibrated radiometer. Record the fixture and detector geometry so a later comparison can reproduce the result.
Reference equipment and accuracy
Use a detector calibrated for the relevant UV band. Spectral mismatch, angle of response and distance contribute to the uncertainty.
When to repeat the check
Review the measurement after lamp replacement or changes to the optics. Operating hours and changes in measured irradiance can be more informative than calendar age.

Technical references Analytik Jena UVP ChemStudio data sheet

Enquire about UV lamp